. 1985 Jul;156(1):203-6.
doi: 10.1148/radiology.156.1.4001406.

Dual-photon Gd-153 Absorptiometry of Bone

Dual-photon Gd-153 Absorptiometry of Bone

H W Wahner et al. Radiology. .

Abstract

Dual-photon absorptiometry with gadolinium 153 was used to measure the mineral content of lumbar vertebrae in cadavers, excised vertebrae with marrow, and dry, marrow-free vertebrae. The error introduced by the surrounding soft tissue of cadavers was 3%, and the error in determining mineral mass or density in excised vertebrae was about 5%. The correlation coefficient between the results of Gd-153 and corrected iodine 125 (single-photon) absorptiometry on 24 femoral necks was 0.99, and the predictive error was 3.7%. Dual-photon absorptiometry accurately indicates bone mass and bone density and is only slightly affected by either surrounding tissue or fat changes in bone marrow.