Evaluation of ion source parameters and liquid chromatography methods for plasma untargeted metabolomics using orbitrap mass spectrometer

J Chromatogr B Analyt Technol Biomed Life Sci. 2025 May 1:1257:124564. doi: 10.1016/j.jchromb.2025.124564. Epub 2025 Mar 19.

Abstract

Although untargeted metabolomics holds promise for study of metabolites in human health and disease, robust method development and optimization are needed to reduce potential analytical biases and to ensure comprehensive, high-throughput results. In this study, the effect of mass spectrometer (MS) ion source parameters on the signal reproducibility and number of metabolite annotations during untargeted metabolomics is shown. Furthermore, different mobile phase gradients and columns (five reversed phase (RP)-C18 and two hydrophilic interaction liquid chromatography (HILIC) columns) were evaluated for untargeted metabolomics of blood plasma extracts. Positioning the electrospray needle at the farthest on the Z-direction and the closest tested position on the Y-direction with respect to the mass spectrometry inlet produced the best signal reproducibility and the greatest number of metabolite annotations. Moreover, optimal ion source conditions included a positive spray voltage between 2.5 and 3.5 kV, a negative spray voltage between 2.5 and 3.0 kV, vaporization and ion transfer tube (ITT) temperature between 250 and 350 °C, 30 to 50 arbitrary units of sheath gas, and at least 10 auxiliary gas units. Despite the differences in chromatographic characteristics, the different RP columns assessed showed comparable performance in terms of number of metabolites annotated. For HILIC columns, a zwitterionic column demonstrated better performance than an amide column. Finally, as compared with use of a RP column alone, use of both the optimal RP and HILIC approaches expanded metabolome coverage: the number of metabolites annotated increased by 60 %. This study highlights the significance of fine-tuning the MS ion source parameters and optimizing chromatographic conditions on metabolome coverage during untargeted metabolomics of plasma samples.

Keywords: ESI needle position; ESI parameters; HILIC columns; Liquid chromatography gradient; Reversed phase columns; Untargeted metabolomics.

MeSH terms

  • Chromatography, Liquid / methods
  • Humans
  • Hydrophobic and Hydrophilic Interactions
  • Mass Spectrometry* / methods
  • Metabolome
  • Metabolomics* / methods
  • Reproducibility of Results