Validation of cobas® pulse point-of-care testing device for blood glucose monitoring

Scand J Clin Lab Invest. 2025;85(4):275-280. doi: 10.1080/00365513.2025.2512382. Epub 2025 May 30.

Abstract

The aim of this study was to validate the blood glucose point-of-care system, cobas® pulse (Roche Diagnostics GmbH), which is the successor to the Accu-Chek® Inform II system (Roche Diagnostics GmbH). Since the cobas® pulse device is intended to replace an existing device from the same manufacturer, we found it highly relevant to perform an industry-independent validation regarding accuracy and comparability with existing glucose measurement systems. From 40 randomly selected, non-fasting adults capillary and venous blood was drawn simultaneously. Correlation and agreement was evaluated by comparing capillary blood glucose on cobas® pulse to plasma glucose on cobas® 8000 (Roche Diagnostics GmbH) and capillary blood glucose on ABL800 Flex (Radiometer, Denmark), respectively. The cobas® pulse generally showed good agreement with both comparison methods, although the agreement between cobas® pulse and ABL800 Flex was better (bias -0.01 mmol/L) than between cobas® pulse and cobas® 8000 (bias 0.61 mmol/L). Differences between measurements of low blood glucose levels (range 0.5 to 4.8 mmol/L) and higher blood glucose levels (range 9.7 to 15.3 mmol/L) when comparing cobas® pulse to ABL800 Flex was also within allowable limits. Altogether, the validation study demonstrated a clinically satisfactory performance of the cobas® pulse point-of-care device.

Keywords: Blood glucose; POCT; clinical chemistry tests; method validation; point-of-care systems; point-of-care testing.

Publication types

  • Validation Study

MeSH terms

  • Adult
  • Blood Glucose Self-Monitoring* / instrumentation
  • Blood Glucose Self-Monitoring* / methods
  • Blood Glucose* / analysis
  • Female
  • Humans
  • Male
  • Middle Aged
  • Point-of-Care Systems*
  • Point-of-Care Testing*
  • Reproducibility of Results

Substances

  • Blood Glucose