[A procedure for the determination of dose performance of radioactive eye applicators (106-Ru/106-Rh) and their significance for indications and treatment in malignant choroid melanomas]

Klin Monbl Augenheilkd. 1985 Nov;187(5):451-4. doi: 10.1055/s-2008-1054371.
[Article in German]

Abstract

Choroidal melanomas occasionally respond very differently to irradiation by ruthenium beta-ray applicators. In order to clarify the role of the radiation parameters on the one hand and the tumor properties on the other, more precise data on the actual surface does rate than those furnished by the supplier are indispensable. With the aid of a measuring method based on the principle of thermoluminescence, the deviation was halved a +/- 15%. The measuring instrument is described, and its impact pact on therapeutic safety and indication limits is discussed.

Publication types

  • English Abstract

MeSH terms

  • Choroid Neoplasms / radiotherapy*
  • Humans
  • Melanoma / radiotherapy*
  • Radioisotopes / therapeutic use
  • Radiotherapy Dosage
  • Rhodium / therapeutic use
  • Ruthenium / therapeutic use
  • Thermoluminescent Dosimetry