Retrieval of occluding unexpanded Palmaz-Schatz stent from a saphenous aorto-coronary vein graft

Cathet Cardiovasc Diagn. 1995 Feb;34(2):159-61. doi: 10.1002/ccd.1810340418.

Abstract

The Palmaz-Schatz coronary stent is used frequently to reduce the rate of restenosis of balloon angioplasty in saphenous vein grafts. In many European centers, the stent, manually crimped on a balloon, is advanced across the stenosis without a protective sheath. This report describes a patient in whom an attempt to deploy a stent in the orifice of a saphenous vein graft was complicated by dislodgement of the unexpanded stent from the balloon. The unexpanded stent caused immediate occlusion of the vein graft and severe ischemic symptoms. The stent was retrieved by inflation of another balloon in the graft, distal to the stent, pulling the balloon toward the guiding catheter, and then withdrawing the whole system.

Publication types

  • Case Reports
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Angina Pectoris / diagnostic imaging
  • Angina Pectoris / therapy
  • Angioplasty, Balloon, Coronary / instrumentation*
  • Coronary Angiography
  • Coronary Artery Bypass*
  • Coronary Disease / diagnostic imaging
  • Coronary Disease / surgery*
  • Equipment Design
  • Equipment Failure
  • Foreign Bodies / diagnostic imaging
  • Foreign Bodies / therapy*
  • Graft Occlusion, Vascular / diagnostic imaging
  • Graft Occlusion, Vascular / therapy*
  • Humans
  • Male
  • Saphenous Vein / transplantation*
  • Stents*