Build-up region of megavoltage photon radiation sources

Med Phys. 1975 Jan-Feb;2(1):14-9. doi: 10.1118/1.594158.

Abstract

Measurements of surface dose and build-up have been measured for photon radiation sources with maximum energy from 1.2 to 25 MeV. A variable volume ionization chamber was used and the results extrapolated to what would be obtained with a zero volume chamber. The results are found to depend systematically on the plate separation of a parallel plate ionization chamber and an empirical method was derived for correcting measurements made with a fixed volume chamber. The relationship of dose build-up curves with skin reactions in radiation therapy patients is discussed.

Publication types

  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Cobalt Radioisotopes
  • Mathematics
  • Physical Phenomena
  • Physics
  • Radiation / instrumentation*
  • Radiation Dosage
  • Radiotherapy, High-Energy

Substances

  • Cobalt Radioisotopes