Use of a new syringe pump (Springfusor) for muscle relaxant infusion

Anaesth Intensive Care. 1993 Aug;21(4):444-6. doi: 10.1177/0310057X9302100414.

Abstract

A new spring-driven syringe pump (Springfusor 10, Go Medical Industries, Subiaco, W.A.) was evaluated for infusion of muscle relaxants in 50 surgical cases. After an initial bolus dose, atracurium was given at a rate of 25 to 37.5 mg/hour and vecuronium at a rate of 4 to 6 mg/hour. The mean infusion rate was found to be 4% above the nominal rate of 5 ml/hour and the accuracy in all cases was within +/- 20%. The Springfusor was found to be rugged, cheap and easy to use. It lacks dose flexibility but is suitable for continuous background infusion of muscle relaxants during surgery.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Atracurium / administration & dosage*
  • Evaluation Studies as Topic
  • Humans
  • Infusion Pumps*
  • Syringes*
  • Vecuronium Bromide / administration & dosage*