[Artifactual defect of inferior myocardium on 123I-metaiodobenzylguanidine myocardial SPECT: characteristic findings and preventive method on phantom study]

Kaku Igaku. 1994 Apr;31(4):359-66.
[Article in Japanese]

Abstract

Discordance between planar and SPECT image of inferior myocardial defect on 123I-metaiodobenzylguanidine myocardial scintigraphy (MIBG) was occasionally observed in the clinical studies. The purpose of this study is to clarify the existence of artifactual myocardial defect on MIBG SPECT image. Phantom study was performed in various kinds of situation using triple gamma camera. Artifactual defect adjacent to the liver was significantly observed in the case of increased liver/heart uptake ratio more than 2:1. Artifactual defect was markedly observed in the 180 degrees than that in the 360 degrees SPECT acquisition. To avoid artifactual defect during SPECT acquisition, liver phantom was moved down to the foot direction (vanishing liver position). In this phantom position, artifactual defect adjacent to the liver was not observed on reconstructed myocardial SPECT image. The vanishing liver position was also applicable to human study. Human body was inclined to right side at an angle of 15 degrees, artifactual defect could be avoided. Further study is needed to analyze the cause of this artifactual defect and to conform the preventive method with vanishing liver position in the routine myocardial SPECT examination.

MeSH terms

  • 3-Iodobenzylguanidine
  • Artifacts*
  • Heart / diagnostic imaging*
  • Humans
  • Iodine Radioisotopes*
  • Iodobenzenes*
  • Liver / diagnostic imaging
  • Models, Structural*
  • Tomography, Emission-Computed, Single-Photon* / methods

Substances

  • Iodine Radioisotopes
  • Iodobenzenes
  • 3-Iodobenzylguanidine