Optimal MR protocol for hepatic hemangiomas. Comparison of conventional spin-echo sequences with T2-weighted turbo spin-echo and serial gradient-echo (FLASH) sequences with gadolinium enhancement

Acta Radiol. 1997 Jul;38(4 Pt 1):565-71. doi: 10.1080/02841859709174388.

Abstract

Purpose: To compare conventional spin-echo (SE) sequences with T2-weighted turbo SE and serial gadolinium-enhanced T1-weighted FLASH sequences in the detection and characterization of hepatic hemangiomas, and to describe the enhancement characteristics of the lesions on dynamic MR imaging.

Material and methods: Forty-two patients with 66 hemangiomas were studied at 1.0 T or 1.5 T by using conventional SE sequences (T1-weighted, T2-weighted, and heavily T2-weighted), T2-weighted turbo SE sequences, and breath-hold T1-weighted FLASH sequences acquired before, immediately after, and 1, 2, 3, 5, and 10 min after injection of a bolus of gadopentetate dimeglumine. Images were quantitatively analyzed for lesion-to-liver contrast-to-noise (C/N) ratios, and qualitatively analyzed for lesion conspicuity. The enhancement pattern and the rapidity of enhancement were analyzed in small (<15 mm), medium (15-39 mm), and large (>39 mm) hemangiomas.

Results: In T2-weighted images, T2-weighted turbo SE and heavily T2-weighted SE images had higher C/N ratios than T2-weighted SE images (p=0.003). Lesion conspicuity was not significantly different in these 3 sequences (p=0.307). In T1-weighted images, T1-weighted FLASH images had higher C/N (p<0.001) and also better lesion conspicuity (p<0.001) than T1-weighted SE images. Immediate uniform enhancement was seen in 43% of small hemangiomas (9 of 21 lesions), and persistent central hypointensity was seen in 73% of large hemangiomas (11 of 15 lesions). Rapid enhancement was seen in 62% of small hemangiomas (13 of 21 lesions) and in 31% of medium or large hemangiomas (14 of 45 lesions).

Conclusion: Further clinical study is needed for evaluating the differential diagnostic advantages of turbo SE T2-weighted imaging compared to the calculation of T2-values by means of a SE T2-weighted sequence. However, the results of the present study suggest that T2-weighted turbo SE imaging and precontrast and serial gadolinium-enhanced FLASH imaging have the potential to replace conventional SE imaging in the evaluation of hepatic hemangiomas.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Contrast Media
  • Diagnosis, Differential
  • Drug Combinations
  • Female
  • Gadolinium
  • Gadolinium DTPA
  • Hemangioma / diagnosis*
  • Humans
  • Liver / pathology*
  • Liver Neoplasms / diagnosis*
  • Magnetic Resonance Imaging / methods*
  • Male
  • Meglumine
  • Middle Aged
  • Organometallic Compounds
  • Pentetic Acid / analogs & derivatives
  • Prospective Studies

Substances

  • Contrast Media
  • Drug Combinations
  • Organometallic Compounds
  • Meglumine
  • Pentetic Acid
  • Gadolinium
  • Gadolinium DTPA