Laboratory testing of the Spiegelberg brain pressure monitor: a technical report

J Neurol Neurosurg Psychiatry. 1997 Dec;63(6):732-5. doi: 10.1136/jnnp.63.6.732.

Abstract

Objectives: The Spiegelberg brain pressure monitor is a low cost intracranial pressure monitoring system that has been used clinically for some time, mainly in Germany. To provide a rigorous bench comparison of the Spiegelberg monitor with the Camino pressure monitor an evaluation programme has been carried out in the UK Shunt Evaluation Laboratory.

Design: Drift over 72 hours and with temperature, a frequency response, and the accuracy of measurement of both static and pulsatile pressures have been tested simultanously in Camino and Spiegelberg transducers using a computerised rig.

Results: Long term zero drift was less than 0.7 mm Hg in both transducers. The Spiegelberg monitor showed no temperature drift whereas the Camino monitor had a drift of around 0.3 mm Hg/degrees C. The Spiegelberg monitor underread mean pressures <40 mm Hg by <1 mm Hg, but the error increased to 4.7 mm Hg at 100 mm Hg. The frequency bandwidth of the Spiegelberg monitor was 4 Hz at a low pressure. Underreading of the amplitude increased with the mean pressure, with a delay of about 0.1 s in the detection of the peaks of pulse waveform.

Conclusion: The Spiegelberg transducer had excellent accuracy for static intracranial pressure measurement, but complex waveform analysis may be biased by its limited dynamic response.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Clinical Laboratory Techniques / instrumentation*
  • Humans
  • Intracranial Pressure*
  • Time Factors
  • Transducers