High-performance liquid chromatographic assay of 8-hydroxyquinoline sulfate and its stability in immunobiological preparations

J Chromatogr A. 1998 Mar 6;798(1-2):173-7. doi: 10.1016/s0021-9673(97)01085-6.

Abstract

Because of the ability of 8-hydroxyquinoline sulfate (8-HQS) to irreversibly bind metals from rubber stoppers, the stability of 8-HQS in tuberculin solutions was investigated. For the determination of 8-HQS, a simple and sensitive reversed-phase HPLC method with detection at 240 nm was developed and validated. Rapid decreases in concentrations of 8-HQS were found in samples stored in original vials which were exposed to different temperatures and vial positions.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Chromatography, High Pressure Liquid / methods*
  • Drug Contamination
  • Drug Stability
  • Drug Storage
  • Kinetics
  • Oxyquinoline / analysis*
  • Temperature
  • Tuberculin / analysis*

Substances

  • Tuberculin
  • Oxyquinoline