Testing criteria for spinal implants: recommendations for the standardization of in vitro stability testing of spinal implants
- PMID: 9629939
- PMCID: PMC3611233
- DOI: 10.1007/s005860050045
Testing criteria for spinal implants: recommendations for the standardization of in vitro stability testing of spinal implants
Abstract
New implants and new surgical approaches should be tested in vitro for primary stability in standardized laboratory tests in order to decide the most appropriate approach before being accepted for clinical use. Due to the complex and still unknown loading of the spine in vivo a variety of different test loading conditions have been used, making comparison of the results from different groups almost impossible. This recommendation was developed in a series of workshops with research scientists, orthopedic and trauma surgeons, and research and development executives from spinal implant companies. The purpose was to agree on in vitro testing conditions that would allow results from various research groups to be compared. This paper describes the recommended loading methods, specimen conditions, and analysis parameters resulting from these workshops.
Similar articles
-
In vitro spinal arthrodesis implant mechanical testing protocols.J Spinal Disord. 1989 Dec;2(4):274-81. J Spinal Disord. 1989. PMID: 2520086
-
Load transfer characteristics between posterior spinal implants and the lumbar spine under anterior shear loading: an in vitro investigation.Spine (Phila Pa 1976). 2012 Aug 15;37(18):E1126-33. doi: 10.1097/BRS.0b013e31825b528d. Spine (Phila Pa 1976). 2012. PMID: 22565384
-
Critical analysis of in vitro stability testing of spinal implants and proposal for standardization.Expert Rev Med Devices. 2022 Mar;19(3):281-286. doi: 10.1080/17434440.2022.2054331. Epub 2022 Mar 21. Expert Rev Med Devices. 2022. PMID: 35297275
-
Practical biomechanical considerations for spine implant testing.Chir Narzadow Ruchu Ortop Pol. 1999;64(3):347-64. Chir Narzadow Ruchu Ortop Pol. 1999. PMID: 10495560 Review.
-
Determination of the in vivo posterior loading environment of the Coflex interlaminar-interspinous implant.Spine J. 2010 Mar;10(3):244-51. doi: 10.1016/j.spinee.2009.10.010. Epub 2009 Dec 11. Spine J. 2010. PMID: 20004622 Review.
Cited by
-
Variability of intervertebral joint stiffness between specimens and spine levels.Front Bioeng Biotechnol. 2024 Feb 29;12:1372088. doi: 10.3389/fbioe.2024.1372088. eCollection 2024. Front Bioeng Biotechnol. 2024. PMID: 38486868 Free PMC article.
-
Lateral fenestration of lumbar intervertebral discs in rabbits: development and characterisation of an in vivo preclinical model with multi-modal endpoint analysis.Eur Spine J. 2024 Feb 19. doi: 10.1007/s00586-024-08153-5. Online ahead of print. Eur Spine J. 2024. PMID: 38372793
-
Application of a combined cancellous lag screw enhances the stability of locking plate fixation of osteoporotic lateral tibial plateau fracture by providing interfragmentary compression force.J Orthop Surg Res. 2024 Feb 14;19(1):139. doi: 10.1186/s13018-024-04564-8. J Orthop Surg Res. 2024. PMID: 38351078 Free PMC article.
-
Cervical spine reconstruction after total vertebrectomy using customized three-dimensional-printed implants in dogs.J Vet Sci. 2024 Jan;25(1):e2. doi: 10.4142/jvs.23241. J Vet Sci. 2024. PMID: 38311317 Free PMC article.
-
Extrusion and meniscal mobility evaluation in case of ramp lesion injury: a biomechanical feasibility study by 7T magnetic resonance imaging and digital volume correlation.Front Bioeng Biotechnol. 2024 Jan 5;11:1289290. doi: 10.3389/fbioe.2023.1289290. eCollection 2023. Front Bioeng Biotechnol. 2024. PMID: 38249805 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources