Comparison of decentration and tilt between one piece and three piece polymethyl methacrylate intraocular lenses

Br J Ophthalmol. 1998 Apr;82(4):419-22. doi: 10.1136/bjo.82.4.419.

Abstract

Background: The extent of the decentration and tilt was prospectively compared between one piece polymethyl methacrylate (PMMA) and three piece PMMA intraocular lenses (IOLs) which were implanted in the capsular bag after performing continuous curvilinear capsulorhexis.

Methods: 91 patients underwent a one piece PMMA IOL implantation in one eye as well as the implantation of the three piece PMMA IOL with polyvinylidene fluoride loops in the opposite eye. The length of the lens decentration and the angle of the tilt were quantitated using the anterior eye segment analysis system (EAS-1000) at 1 week as well as 1, 3, and 6 months postoperatively.

Results: The mean length of the decentration in the one piece IOL was smaller than that in the three piece IOL at 1 week (p = 0.0092), 1 month (p = 0.0044), 3 months (p = 0.0069), and 6 months (p = 0.0010) postoperatively. However, no significant difference was found in the degree of the tilt between the two types of IOLs throughout the observation periods.

Conclusion: These results clarified that the one piece PMMA IOL with rigid PMMA haptics implanted in the capsular bag provides a better centration than the three piece PMMA IOL with flexible haptics, whereas the tilt was the same between the two types of IOLs.

Publication types

  • Comparative Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Female
  • Humans
  • Lens Implantation, Intraocular*
  • Lenses, Intraocular*
  • Male
  • Middle Aged
  • Ophthalmology*
  • Polymethyl Methacrylate
  • Polyvinyls
  • Prospective Studies

Substances

  • Polyvinyls
  • polyvinylidene fluoride
  • Polymethyl Methacrylate