. 1998 Apr-May;19(7-9):643-9.
doi: 10.1016/s0142-9612(97)00158-0.

Microscopical features in retrieved human Branemark implants: a report of 19 cases

Affiliations

Microscopical features in retrieved human Branemark implants: a report of 19 cases

A Piattelli et al. Biomaterials. 1998 Apr-May.

Abstract

The authors present a histologic analysis of 19 Branemark titanium implants retrieved for different causes: four implants were removed for abutment fracture, one for dental nerve dysesthesia, two for bone overheating, two for peri-implantitis, nine for mobility, one for unknown causes. In the implants removed for fracture a high bone-implant contact percentage was present (71.83 +/- 4.96%) with compact, mature bone at the interface. The picture of the failure due to bone overheating was characteristic with the presence of bone sequestra and of a gap between implant and bone filled by lymphocytes and plasma cells: many bacteria surrounded the necrotic bone and no newly regenerated bone was present. In peri-implantitis an inflammatory infiltrate was observed in the peri-implant tissues: a dense fibrous connective tissue was present around implants failed for mobility. The microscopical picture is certainly extremely important in identifying the causal determinants of an implant failure.

PubMed Disclaimer

Similar articles

Cited by

Publication types

LinkOut - more resources