[Use of a multileaf collimator for the production of intensity-modulated beams]

Cancer Radiother. 1998 Jul-Aug;2(4):392-403. doi: 10.1016/s1278-3218(98)80352-1.
[Article in French]

Abstract

In external radiotherapy, the use of intensity modulated fields has been proposed for tissue and non-homogeneity compensation or for the generation of conformal dose distributions. Multileaf collimators can be employed dynamically for the modulation of the X-ray field in two dimensions. Efficient dynamic collimation became possible due to advances in computer and linear accelerator technology. It presents a number of advantages over conventional methods such as the use of compensators. We have developed a program which calculates, from a given intensity distribution, the motion of the MLC leaves as a function of monitor units, and we have applied it on a Varian linear accelerator with a 40 pair multileaf collimator. The analysis of the experimental results demonstrates the feasibility and the potential of the method.

Publication types

  • English Abstract

MeSH terms

  • Algorithms*
  • Equipment Design
  • Feasibility Studies
  • Humans
  • Image Processing, Computer-Assisted
  • Movement
  • Particle Accelerators
  • Phantoms, Imaging
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted / instrumentation*
  • Tomography, X-Ray Computed