Rapid MRI and velocimetry of cylindrical Couette flow

Magn Reson Imaging. 1998 Oct;16(8):953-61. doi: 10.1016/s0730-725x(98)00089-7.

Abstract

A narrow-gap, temperature-controlled Couette flow rheometer has been developed to study fluid velocities within the annular gap between two concentric cylinders by nuclear magnetic resonance (NMR) imaging and velocimetry. Alternative pulsed-field-gradient-based nuclear magnetic resonance imaging strategies which may be used for measurement of velocity within the Couette flow device have been evaluated. These include two-dimensional (2-D) imaging techniques with acquisition times of several minutes and a one-dimensional (1-D) projection method which exploits the symmetry of the device to reduce overall measurement time to less than 1 min. Velocity measurements made using each technique are presented for a Newtonian fluid undergoing Couette flow at shear rates of approximately 20 and 60 s(-1).

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Blood Flow Velocity
  • Humans
  • Image Processing, Computer-Assisted
  • Magnetic Resonance Imaging* / instrumentation
  • Magnetic Resonance Imaging* / methods
  • Models, Cardiovascular
  • Rheology / instrumentation