Endovascular baroreflex amplification for resistant hypertension: a safety and proof-of-principle clinical study

Lancet. 2017 Dec 16;390(10113):2655-2661. doi: 10.1016/S0140-6736(17)32337-1. Epub 2017 Sep 1.

Abstract

Background: Carotid baroreflex activation lowers blood pressure and might have potential application for the treatment of resistant hypertension. We did a proof-of-principle trial with a novel endovascular baroreceptor amplification device, MobiusHD (Vascular Dynamics, Mountain View, CA, USA), in patients with resistant hypertension.

Methods: CALM-FIM_EUR was a prospective, first-in-human, open-label study done at six European centres. Eligible patients were adults with resistant hypertension (office systolic blood pressure ≥160 mm Hg despite taking at least three antihypertensive agents, including a diuretic). MobiusHD devices were implanted unilaterally in the internal carotid artery. The primary endpoint was the incidence of serious adverse events at 6 months. Secondary endpoints included changes in office and 24 h ambulatory blood pressure. This trial is registered with ClinicalTrials.gov, number NCT01911897.

Findings: Between December, 2013, and February, 2016, 30 patients were enrolled and underwent successful implantation. Mean age was 52 years (SD 12), 15 patients (50%) were men, and mean antihypertensive use was 4·4 drugs (1·4). Mean office blood pressure was 184/109 mm Hg (18/14) at baseline and was reduced by 24/12 mm Hg (13-34/6-18) at 6 months (p=0·0003 for systolic and p=0·0001 diastolic blood pressure). Mean baseline 24 h ambulatory blood pressure was 166/100 mm Hg (17/14) at baseline and was reduced by 21/12 mm Hg (14-29/7-16) at 6 months (p<0·0001 for systolic and diastolic blood pressure). Five serious adverse events had occurred in four patients (13%) at 6 months: hypotension (n=2), worsening hypertension (n=1), intermittent claudication (n=1) and wound infection (n=1).

Interpretation: In patients with resistant hypertension, endovascular baroreceptor amplification with the MobiusHD device substantially lowered blood pressure with an acceptable safety profile. Randomised, double-blind, sham-controlled trials are warranted to investigate the use of this treatment further.

Funding: Vascular Dynamics.

Publication types

  • Clinical Trial

MeSH terms

  • Baroreflex / physiology*
  • Blood Vessel Prosthesis Implantation*
  • Blood Vessel Prosthesis*
  • Carotid Sinus / physiology*
  • Female
  • Humans
  • Hypertension / therapy*
  • Male
  • Middle Aged
  • Prospective Studies

Associated data