Multicenter evaluation of workflow performances of the new Olympus AU3000i

Clin Lab. 2009;55(11-12):459-64.

Abstract

Background: A new immunoanalyzer (AU3000i) has been developed by Olympus (Rungis, France) with the intention of offering a consolidated workcell. The aim of this experiment was to assess, in a multi-centre study (two French sites, one German site), the practicability of the Olympus AU3000i in terms of throughput (test/h), rerun capabilities, emergency sample handling and reflex test capabilities.

Results: The workload study showed that the AU3000i was capable of running both one- and two-step assay protocols with a throughput of 240 tests per hour which corresponds to 209.3 tests per hour including the initialization time. Emergency samples were correctly treated as a priority in less than 30 minutes for ten samples. Furthermore the analyzer could be programmed to generate automatic dilution, rerun and reflex tests, reducing the hands-on labour time for technicians.

Conclusion: Workflow studies showed that Olympus AU3000i and AU Clinical Chemistry can cover the combined workload of various routine analyzers in a variety of laboratory environments.

Publication types

  • Evaluation Study
  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • High-Throughput Screening Assays / instrumentation*
  • Humans
  • Immunoassay / instrumentation*
  • Laboratories, Hospital