Factory-Calibrated Continuous Glucose Monitoring: How and Why It Works, and the Dangers of Reuse Beyond Approved Duration of Wear

Diabetes Technol Ther. 2019 Apr;21(4):222-229. doi: 10.1089/dia.2018.0401. Epub 2019 Feb 28.

Abstract

Continuous glucose monitors (CGM) display real-time glucose values enabling greater glycemic awareness with reduced management burden. Factory-calibrated CGM systems allow for glycemic assessment without the pain and inconvenience of fingerstick glucose testing. Advances in sensor chemistry and CGM algorithms have enabled factory-calibrated systems to have greater accuracy than previous generations of CGM technology. Despite these advances many patients and providers are hesitant about the idea of removing fingerstick testing from their diabetes care. In this commentary, we aim to review the clinical trials on factory-calibrated CGM systems, present the algorithms which facilitate factory-calibrated CGMs to improve accuracy, discuss clinical use of factory-calibrated CGMs, and finally present two cases demonstrating the dangers of utilizing exploits in commercial systems to prolong sensor life.

Keywords: Continuous glucose monitoring; Factory calibration; Type 1 diabetes.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Blood Glucose / analysis*
  • Blood Glucose Self-Monitoring*
  • Calibration
  • Diabetes Mellitus, Type 1 / blood*
  • Humans

Substances

  • Blood Glucose