Simultaneous determination of estriol and estriol 3-sulfate in serum by column-switching semi-micro high-performance liquid chromatography with ultraviolet and electrochemical detection

J Chromatogr B Biomed Sci Appl. 1999 Feb 19;723(1-2):39-45. doi: 10.1016/s0378-4347(98)00555-6.

Abstract

A column-switching HPLC with semi-microcolumn enabled us a direct and simultaneous analysis of estriol (E3) and estriol 3-sulfate (E3 S) in human serum in combination with ultraviolet (for E3 S) and electrochemical (for E3) detectors. The mobile phases (phosphate buffer pH 7.0) contained 5 mM tetra-n-butylammonium ion (TBA) as a counter ion for E3 S. Serum samples were diluted with 200 mM phosphate buffer (pH 7.0) containing 100 mM TBA, then injected to the pre-column. After serum proteins had flowed out from the pre-column, E3 and E3 S were transferred to the enrichment column. Subsequently the analytes were eluted to the analytical column. Detection limits of E3 and E3 S in human serum were 2.5 ng/ml and 295 ng/ml. Serum E3 and E3 S levels (mean +/- SD) of umbilical artery from 18 full-term healthy neonates were 33+/-23 ng/ml and 1.26+/-0.69 microg/ml, respectively.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Calibration
  • Chromatography, High Pressure Liquid / methods*
  • Electrochemistry
  • Estriol / analogs & derivatives*
  • Estriol / blood*
  • Humans
  • Infant, Newborn
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Spectrophotometry, Ultraviolet

Substances

  • estriol 3-sulfate
  • Estriol