Verification of improved patient outcomes with a partially implantable hearing aid, The SOUNDTEC direct hearing system

Laryngoscope. 2001 Oct;111(10):1682-6. doi: 10.1097/00005537-200110000-00002.

Abstract

Introduction: Partially implantable hearing devices have been developed to address some of the user-perceived shortcomings of standard amplification systems. Partially implantable devices are purported to provide improved sound quality as a result of decreased occlusion, decreased feedback, and enhanced clarity resulting from increased high-frequency gain. Such improvements may result in greater user satisfaction. To justify selection of a partially implantable device and undergoing a minor surgical procedure, verification techniques must be used to document user improvement or increased satisfaction over conventional amplification.

Objective: To evaluate patient satisfaction with the SOUNDTEC direct hearing system.

Study design: Within-subjects repeated measures design.

Methods: Objective and subjective evaluation pre- and post-implantation with the SOUNDTEC device. Verification techniques included tonal functional gain measures with traditional amplification and the SOUNDTEC device, word recognition in quiet (NU-6) and in noise (SPIN), the Abbreviated Profile of Hearing Aid Benefit (APHAB), and the Hough Ear Institute Profile (HEIP).

Results: Although there was no significant difference between optimal traditional amplification and the SOUNDTEC device for speech perception measures, the SOUNDTEC device yielded statistically significant increased high-frequency functional gain. Subjective reports indicated that the SOUNDTEC device provides a cleaner, more natural sound without feedback than traditional amplification.

Conclusions: Partially implantable hearing aids may address some of the limitations of traditional amplification systems.

Publication types

  • Clinical Trial
  • Evaluation Study

MeSH terms

  • Adult
  • Aged
  • Audiometry, Pure-Tone
  • Female
  • Hearing Aids*
  • Hearing Loss, Sensorineural / rehabilitation*
  • Humans
  • Male
  • Middle Aged
  • Patient Satisfaction
  • Prosthesis Design
  • Prosthesis Implantation / instrumentation*
  • Speech Discrimination Tests