Simplified high-performance liquid chromatographic method for determination of risperidone and 9-hydroxyrisperidone in plasma after overdose

J Chromatogr B Analyt Technol Biomed Life Sci. 2002 Jun 5;772(2):373-8. doi: 10.1016/s1570-0232(02)00091-0.

Abstract

For toxicological purposes, a HPLC assay was developed for the simultaneous determination of risperidone and 9-hydroxyrisperidone in human plasma. After a single-step liquid-liquid extraction, both compounds were separated on a C(18) column and measured at 280 nm. A good inter-assay accuracy (<116%) was achieved with inter-assay precision less than 12%. Quantification limits were 10 ng/ml. This rapid method (run time <5 min) is currently used for poison management.

MeSH terms

  • Antipsychotic Agents / adverse effects
  • Antipsychotic Agents / blood*
  • Chromatography, High Pressure Liquid / methods*
  • Drug Overdose / blood*
  • Humans
  • Isoxazoles / blood*
  • Paliperidone Palmitate
  • Pyrimidines / blood*
  • Reproducibility of Results
  • Risperidone / adverse effects
  • Risperidone / blood*
  • Sensitivity and Specificity

Substances

  • Antipsychotic Agents
  • Isoxazoles
  • Pyrimidines
  • Risperidone
  • Paliperidone Palmitate