Clinical course and management strategies for hemolysis after transcatheter closure of patent arterial ducts

Catheter Cardiovasc Interv. 2003 Aug;59(4):538-43. doi: 10.1002/ccd.10593.

Abstract

Residual flows following transcatheter coil or device closure of the patent ductus arteriosus (PDA) can result in hemolysis. Of 611 patients who underwent transcatheter PDA closure at our institution, 5 patients (age, 6-63 years) developed overt hemolysis (after coil occlusion in 4 and Amplazter device closure in 1). All had ducts > 3 mm and residual flows after the procedure. In one patient, hemolysis occurred 3 months after coil occlusion following a period of uncontrolled hypertension. The occurrence of hemolysis correlated significantly with both age as well as duct size (P < 0.00001). Hemolysis was associated with a fall in hemoglobin of 3-6 g/100 ml (n = 3), jaundice (n = 2), and renal failure (n = 1). Hemolysis subsided spontaneously in one patient and four patients required flow elimination. Deploying additional coils in three patients eliminated residual flows. In one patient (after Amplatzer device closure for 12.5 mm duct with aneurysm), flow persisted after 25 additional coils, transient balloon occlusion, and gel foam instillation. Flow elimination was eventually achieved through thrombin instillation after balloon occlusion of the ampulla. All patients recovered completely and were well on follow-up. Although hemolysis after duct occlusion is rare (0.8% in this series), residual flow at the end of the procedure merits careful monitoring. Aggressive elimination of residual flows is often necessary to control hemolysis.

Publication types

  • Review

MeSH terms

  • Adult
  • Anemia, Hemolytic / etiology*
  • Anemia, Hemolytic / therapy*
  • Cardiac Catheterization / adverse effects*
  • Child
  • Ductus Arteriosus, Patent / therapy*
  • Embolization, Therapeutic / adverse effects*
  • Female
  • Humans
  • Middle Aged