Low-risk palpable breast masses removed using a vacuum-assisted hand-held device

Am J Surg. 2003 Oct;186(4):362-7. doi: 10.1016/s0002-9610(03)00263-0.

Abstract

Background: This study evaluates the safety, efficacy, and patient acceptance of a vacuum-assisted, hand-held biopsy device (Mammatome) in the percutaneous removal of breast masses using ultrasound guidance.

Methods: A multicenter, nonrandomized study evaluated 216 women with low-risk palpable lesions. Lesions 1.5 to 3.0 cm in size were removed using an 8-gauge probe. Those lesions <1.5 cm were removed with the 11-gauge probe. Follow-up evaluation was performed at 10 days and 6 months after biopsy.

Results: A total of 127 patients had biopsies using the 8-gauge probe, and 89 patients had biopsies using the 11-gauge probe. At 6-month follow-up, 98% of the lesions remained nonpalpable, 73% with no ultrasonographically visible evidence of the original lesion. Most complications were mild and anticipated. Most patients (98%) were satisfied with incision appearance, and 92% of patients would recommend the procedure to others.

Conclusions: Percutaneous removal of palpable benign breast masses using the Mammotome system is feasible and safe, and yields high patient satisfaction. The results at 6 months after biopsy demonstrated the effectiveness of benign lesion removal, with correlative clinical data demonstrating lack of palpability and no need for additional procedures. Continuing evaluation of long-term efficacy is ongoing.

Publication types

  • Clinical Trial
  • Multicenter Study
  • Research Support, Non-U.S. Gov't