Measurement of beam-axis displacement from the isocenter during three-dimensional conformal radiosurgery with a micro-multileaf collimator

Radiother Oncol. 2004 Jan;70(1):45-8. doi: 10.1016/j.radonc.2003.11.008.

Abstract

We describe the displacement of the beam-axis from the planning isocenter in clinical situations during three-dimensional conformal radiosurgery using an Acculeaf bi-directional micro-multileaf collimator. The displacements were recorded for 64 ports using a video imaging system and a stereotactic arc. The mean displacement was 0.41+/-0.25 mm.

Publication types

  • Evaluation Study

MeSH terms

  • Brain Neoplasms / surgery
  • Humans
  • Imaging, Three-Dimensional
  • Radiosurgery / instrumentation*
  • Radiotherapy, Conformal / instrumentation
  • Radiotherapy, Conformal / methods*
  • Stereotaxic Techniques