High-performance liquid chromatographic assay of clonazepam in human plasma using a non-porous silica column

Biol Pharm Bull. 2004 Jun;27(6):893-5. doi: 10.1248/bpb.27.893.

Abstract

A new HPLC method has been developed for measuring clonazepam (CZP) in plasma, using a reversed-phase non-porous silica column packed with 2 microm particles. CZP in plasma was first purified with a column extraction technique and injected onto a non-porous silica column. The calibration curve was linear from 5-200 ng/ml. The recoveries of CZP added to plasma were more than 94.0%, with a coefficient of variation in the range of 5.1-13.8%. We developed a rapid routine method using a non-porous silica column that was accurate and improved solvent consumption in the measurement of CZP.

Publication types

  • Comparative Study

MeSH terms

  • Chromatography, High Pressure Liquid / methods
  • Clonazepam / blood*
  • Humans
  • Sensitivity and Specificity
  • Silicon Dioxide / analysis*

Substances

  • Clonazepam
  • Silicon Dioxide