Validation of a liquid chromatography assay for the quantification of the Raf kinase inhibitor BAY 43-9006 in small volumes of mouse serum

J Chromatogr B Analyt Technol Biomed Life Sci. 2004 Sep 25;809(1):99-103. doi: 10.1016/j.jchromb.2004.06.003.

Abstract

BAY 43-9006 is a selective Raf-1 kinase inhibitor with antitumor activity against a variety of human cancers. A highly sensitive HPLC method for determination of BAY 43-9006 in small volumes of serum (30 microl) was developed. Sample preparation involved a liquid-liquid extraction procedure with tolnaftate as internal standard followed by linear gradient elution at a reversed phase C18 column and UV detection. The method was selective and the calibration curves were linear over the concentration range of 80-2000 ng/ml. The intra-day accuracy ranged from 99.9 to 107.6% and the inter-day accuracy from 94.6 to 115%. The lower limit of quantitation (LOQ) was 80 ng/ml with an accuracy of 105.8%. Thus, this method has been validated and can be applied for the drug monitoring or pharmacokinetic studies of BAY 43-9006 in small volumes of serum samples.

Publication types

  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Animals
  • Benzenesulfonates / blood*
  • Calibration
  • Enzyme Inhibitors / blood*
  • Mice
  • Niacinamide / analogs & derivatives
  • Phenylurea Compounds
  • Proto-Oncogene Proteins c-raf / antagonists & inhibitors*
  • Pyridines / blood*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Sorafenib

Substances

  • Benzenesulfonates
  • Enzyme Inhibitors
  • Phenylurea Compounds
  • Pyridines
  • Niacinamide
  • Sorafenib
  • Proto-Oncogene Proteins c-raf