Time-resolved undersampled projection reconstruction magnetic resonance imaging of the peripheral vessels using multi-echo acquisition

Magn Reson Med. 2005 Mar;53(3):730-4. doi: 10.1002/mrm.20404.

Abstract

The hybrid projection reconstruction (PR) imaging provides high temporal resolution through an undersampled PR acquisition for the in-plane dimensions and Cartesian slice encoding for the through-plane dimension. The undersampling of projection data introduces streak artifact, which may severely compromise image quality. This study reports on a combination of multi-echo acquisition with time-resolved undersampled PR imaging and its application to peripheral magnetic resonance angiography. Multi-echo acquisition improved imaging speed effectively, thereby reducing the undersampling streak artifact and improving the temporal resolution. The gradient distortion was reduced through gradient calibration and accurate k-space trajectory measurement.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Artifacts
  • Contrast Media
  • Gadolinium DTPA
  • Humans
  • Image Processing, Computer-Assisted
  • Leg / blood supply
  • Magnetic Resonance Angiography / methods*
  • Phantoms, Imaging
  • Time Factors