Native aortic valve insufficiency in patients with left ventricular assist devices

Ann Thorac Surg. 2006 Feb;81(2):e6-8. doi: 10.1016/j.athoracsur.2005.08.072.

Abstract

Management of mild to moderate aortic insufficiency in patients with a left ventricular assist device remains controversial. We report 3 patients with aortic insufficiency and pulsatile left ventricular assist devices treated with a central aortic valve coapting suture. Two of the repairs have been durable for more than 1 year and aspirin appears to be sufficient anticoagulation.

Publication types

  • Case Reports

MeSH terms

  • Aged
  • Aortic Valve Insufficiency / surgery*
  • Cardiac Surgical Procedures / methods*
  • Heart-Assist Devices*
  • Humans
  • Male
  • Middle Aged
  • Suture Techniques
  • Treatment Outcome