An electrochemical biosensor using tyrosinase was constructed for the determination of catechol. The enzyme was extracted from a plant source Amorphophallus companulatus and entrapped in agarose-guar gum composite biopolymer matrix. Catechol was determined by direct reduction of biocatalytically liberated quinone species at -0.1 V versus Ag/AgCl (3M KCl). The response was found to be linear and concentration dependent in the range of 6 x 10(-5) to 8 x 10(-4)M with a lower detection limit of 6 microM. It has reusability up to 20 cycles and a shelf life of more than 2 months when stored at 4 degrees C.