Balloon-expandable stenting with and without coiling for wide-neck and complex aneurysms

Surg Neurol. 2006 Dec;66(6):603-10; discussion 610. doi: 10.1016/j.surneu.2006.05.058.

Abstract

Background: Wide-necked, saccular, dissecting, and fusiform intracranial aneurysms are poor coil retainers. Retention can be improved by parent-artery stenting across the aneurysm.

Methods: We used a balloon-expandable stent and delivery system, intending to treat 38 aneurysms in 36 patients. Stents could not be advanced across the neck of 2 aneurysms near the ophthalmic artery origin. These cases were managed by temporary balloon remodeling and coiling. Stenting alone was done for 15 aneurysms, including 7 in vertebral artery V4 segments. Stenting with immediate or delayed coiling was done in 21 aneurysms.

Results: Stenting alone caused immediate and complete obliteration of 1 treated aneurysm (7%), subtotal obliteration in 13 treated (86%) aneurysms, and was associated with 1 failure. Stenting and coiling yielded a significantly better 57% complete obliteration rate, 43% subtotal obliteration, and no failures. There were 5 complications: 1 wire perforation, 2 cavernous-carotid-sinus fistulae, and 2 partial in-stent thromboses. All were controlled or cleared with no long-term sequelae or deaths. Contrast imaging at 1 to 12 months was available for 30 patients (13 stent-only, 17 stent-plus-coiling), demonstrating complete obliteration in 25 (83%) and subtotal obliteration in 5. A total of 7 stent-only aneurysms (4 V4s) were completely obliterated, and 3 (all V4s) were > or = 90% obliterated.

Conclusion: Stenting and coiling through the wall of the stent resulted in 88% (15/17) complete obliteration when imaged 1 to 12 months after treatment. Stenting alone effectively closed off V4-segment wide-necked aneurysms but was inferior to stenting and coiling in less mobile vessels.

Publication types

  • Multicenter Study

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Angioplasty, Balloon / instrumentation*
  • Angioplasty, Balloon / methods*
  • Basilar Artery / diagnostic imaging*
  • Basilar Artery / surgery*
  • Cerebrovascular Disorders / diagnostic imaging
  • Cerebrovascular Disorders / epidemiology
  • Cerebrovascular Disorders / etiology
  • Equipment Design
  • Equipment Failure
  • Female
  • Humans
  • Intracranial Aneurysm / diagnostic imaging*
  • Intracranial Aneurysm / surgery*
  • Intraoperative Complications / epidemiology
  • Male
  • Middle Aged
  • Monitoring, Intraoperative
  • Neurosurgical Procedures / instrumentation
  • Radiography
  • Stents*
  • Vertebral Artery / diagnostic imaging*
  • Vertebral Artery / surgery*