Focused intracochlear electric stimulation with phased array channels

J Acoust Soc Am. 2007 Jun;121(6):3703-16. doi: 10.1121/1.2722047.

Abstract

A method is described for producing focused intracochlear electric stimulation using an array of N electrodes. For each electrode site, N weights are computed that define the ratios of positive and negative electrode currents required to produce cancellation of the voltage within scala tympani at all of the N-1 other sites. Multiple sites can be stimulated simultaneously by superposition of their respective current vectors. The method allows N independent stimulus waveforms to be delivered to each of the N electrode sites without spatial overlap. Channel interaction from current spread associated with monopolar stimulation is substantially eliminated. The method operates by inverting the spread functions of individual monopoles as measured with the other electrodes. The method was implemented and validated with data from three human subjects implanted with 22-electrode perimodiolar arrays. Results indicate that (1) focusing is realizable with realistic precision; (2) focusing comes at the cost of increased total stimulation current; (3) uncanceled voltages that arise beyond the ends of the array are weak except when stimulating the two end channels; and (4) close perimodiolar positioning of the electrodes may be important for minimizing stimulation current and sensitivity to measurement errors.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Cochlea / physiology*
  • Cochlear Implants*
  • Electric Stimulation
  • Electrodes
  • Equipment Design
  • Humans
  • Mathematics
  • Models, Theoretical