Agreement between optical coherence tomography and digital stereophotography in vertical cup-to-disc ratio measurement

Graefes Arch Clin Exp Ophthalmol. 2009 Mar;247(3):377-83. doi: 10.1007/s00417-008-0968-3. Epub 2008 Oct 29.

Abstract

Background: Given the increasing popularity of optical coherence tomography in the diagnosis of glaucoma, we aimed to assess agreement between StratusOCT and stereophotography in measuring the vertical cup-to-disc ratio (VCDR) and examine whether it can be improved by changing the StratusOCT cup-offset.

Methods: In group 1 (79 eyes), VCDR measurements obtained by StratusOCT (using the default cup-offset of 150 microm above the retinal pigment epithelium) and stereophotography were compared. For each eye, the cup-offset was moved upward or downward until the difference disappeared and an adjusted cup-offset was identified. In group 2 (41 eyes), the stereophotography VCDR was compared to the StratusOCT value obtained by using the mean adjusted cup-offset. StratusOCT was used to classify optic disc cups as shallow or deep.

Results: In group 1, StratusOCT overestimated VCDR in optic discs with deep cups and underestimated it in those with shallow cups (p < 0.0001). Mean adjusted cup-offsets were 72.8 +/- 92 microm and 191.6 +/- 70 microm for eyes with deep and shallow cups. The 95% limits of agreement (LOA) ranged from -0.21 to 0.17. In group 2, the adjusted cup-offsets enabled us to eliminate the statistical differences in VCDR measurements and improve the 95% LOA (from -0.13 to 0.14).

Conclusion: Adjusting the StratusOCT cup-offset according to the cup depth improves agreement between StratusOCT and stereophotography in measuring the VCDR.

Publication types

  • Comparative Study

MeSH terms

  • Diagnostic Techniques, Ophthalmological*
  • Female
  • Glaucoma, Open-Angle / diagnosis*
  • Humans
  • Male
  • Middle Aged
  • Optic Disk / pathology*
  • Optic Nerve / pathology*
  • Photography / methods*
  • Reproducibility of Results
  • Tomography, Optical Coherence / methods*