[Transcatheter pulmonary valve replacement in sheep: 6-month evaluation of self-expanding valved stents]

Zhonghua Xin Xue Guan Bing Za Zhi. 2008 Feb;36(2):151-5.
[Article in Chinese]

Abstract

Objective: The purpose of this study was to determine the feasibility of transcatheter pulmonary valve replacement in sheep up to 6 months post procedure.

Methods: Fresh sheep pericardium treated with a 0.6% glutaraldehyde solution for 36 hours was sutured to a valvular ring and then fixed onto a newly designed nitinol self-expandable stent. Thoracotomy was performed in sheep (23.5 +/- 3.1) kg under general anesthesia and the device was delivered into the native pulmonary valve of the sheep via the anterior wall of right ventricle by catheter and fooled for 6 months.

Results: One sheep died 4 months after the procedure due to in-stent thrombosis. Another 4 animals survived the 6-month observing period. Angiographic and hemodynamic measurements confirmed good positioning and function of the stents with a competent valve immediately post procedure and 6 months post the procedure in surviving animals.

Conclusion: Implantation of the nitinol self-expandable stent in the pulmonary valve position by a transcatheter approach is feasible and good function of transcatheter implanted memory nitinol valved stents was shown after 6 months of implantation in sheep.

Publication types

  • English Abstract
  • Research Support, Non-U.S. Gov't