Laser-assisted extraction of pacemaker and defibrillator leads: the role of the cardiac surgeon

Ann Thorac Surg. 2009 May;87(5):1446-50; discussion 1450-1. doi: 10.1016/j.athoracsur.2009.02.015.

Abstract

Background: The development of laser-assisted extraction techniques for chronically implanted pacemaker and defibrillator leads has reduced the need for open surgical removal. Reports of the mortality from laser-assisted extraction range from 1.9% to 3.4%. The purpose of this study was to determine the rate of major cardiovascular injury and emphasize the need for cardiothoracic surgical participation in this procedure.

Methods: A retrospective cohort study was performed of 112 consecutive laser-assisted lead extractions at a single university medical center during a 6-year period. Patient and lead characteristics were analyzed as well as indications, outcomes, and major complications.

Results: Successful lead extraction was accomplished in 103 (92%) of the 112 patients. Elective sternotomy after failure of laser-assisted lead removal was successfully performed in 4 patients. Emergent surgical intervention was required in 4 patients for caval perforation (n = 2), subclavian vein injury (n = 1), or right atrial injury (n = 1). Three of the 4 patients requiring emergent intervention died, for an overall series mortality of 2.6%. In July of 2006, a policy of cardiothoracic surgeon presence during the laser-assisted extraction was instituted. Since that time, there has been one emergent sternotomy and one elective sternotomy for lead removal with no procedure-related deaths.

Conclusions: Despite recent advances in laser technology for the removal of pacemaker and defibrillator leads, the potential for major cardiovascular injury and death remains. Involvement of the cardiothoracic surgeon in both the preoperative decision-making process as well as the laser-assisted lead extraction is critical to prevent or emergently treat any major complications.

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Cardiac Surgical Procedures / standards*
  • Defibrillators, Implantable / adverse effects*
  • Device Removal / instrumentation
  • Device Removal / methods*
  • Electrodes, Implanted / adverse effects*
  • Equipment Failure
  • Female
  • Heart Arrest / etiology
  • Humans
  • Lasers
  • Male
  • Middle Aged
  • Pacemaker, Artificial / adverse effects*
  • Postoperative Complications / epidemiology*
  • Retrospective Studies
  • Sick Sinus Syndrome / etiology
  • Tachycardia, Ventricular / etiology
  • Thoracic Surgery / statistics & numerical data*