The effects of closed endotracheal suction on ventilation during conventional and high-frequency oscillatory ventilation

Pediatr Res. 2009 Oct;66(4):400-4. doi: 10.1203/PDR.0b013e3181b337ec.

Abstract

In newborn infants, closed endotracheal tube (ETT) suction may reduce associated adverse effects, but it is not clear whether ventilation is maintained during the procedure. We aimed to determine the effect of ETT size, catheter size, and suction pressure on ventilation parameters measured distal to the ETT. Suction was performed on a test lung, ventilated with conventional (CMV) and high-frequency oscillatory ventilation (HFOV) using ETT sizes 2.5-4.0 mm, catheter sizes 5-8 French gauge (Fr), and suction pressures 80-200 mm Hg. Tracheal and circuit peak inspiratory pressure, positive end-expiratory pressure, and tracheal tidal volume (VT) were recorded for each suction episode. During both CMV and HFOV, tracheal pressures and VT were considerably reduced by suctioning; this reduction was dependent on the combination of ETT, catheter, and suction pressure. Loss of VT, inflation pressure (CMV), and pressure amplitude (HFOV) occurred primarily with insertion of the catheter, and loss of end-expiratory pressure (CMV) and mean tracheal pressure (HFOV) occurred with the application of suction. Circuit pressures were reduced to lesser degree. We conclude that airway pressures and VT are not maintained during closed endotracheal suction with either CMV or HFOV, and choice of equipment and settings will affect the degree of interruption to ventilation.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • High-Frequency Ventilation* / methods
  • Humans
  • Infant, Newborn
  • Intubation, Intratracheal* / adverse effects
  • Intubation, Intratracheal* / instrumentation
  • Intubation, Intratracheal* / methods
  • Positive-Pressure Respiration
  • Suction* / adverse effects
  • Suction* / instrumentation
  • Suction* / methods