Attenuation correction of somatostatin receptor SPECT by integrated low-dose CT: is there an impact on sensitivity?

Clin Nucl Med. 2009 Dec;34(12):869-73. doi: 10.1097/RLU.0b013e3181becfcb.

Abstract

Aim: Somatostatin receptor scintigraphy (SRS) is an established imaging modality for neuroendocrine tumors (NET). Additional single photon emission computed tomography (SPECT-CT) not only permits image fusion but also attenuation correction (AC) of SPECT data. This study evaluated whether attenuation corrected SPECT-images (SPECT[AC]) are more sensitive than nonattenuation corrected SPECT-reconstructions (SPECT[NAC]) for the detection of NET lesions.

Methods: The imaging data (planar In-111-octreotide scintigraphy and SPECT-CT) of 50 consecutive patients (28 male; 22 female; age, 34-80; mean, 65 years) with NET were included in this retrospective analysis. SPECT data were reconstructed with and without integrated CT-based AC and then analyzed by 2 experienced readers for the presence of pathologic uptake in a blinded consensus reading. Fused SPECT-CT, contemporary CT/MRI, and clinical as well as imaging follow-up served as a reference standard. All foci were rated in both the SPECT(NAC)- and SPECT(AC)-reconstructions for intensity and contrast using a 6-point-score ("0 = no uptake/no delineation from surrounding tissue" to "5 = very high uptake/very strong delineation from surrounding tissue"). The scores were analyzed in a 6 x 6 contingency table using the McNemar Bowker test.

Results: A total of 222 pathologic foci were detected by SPECT(NAC) and 227 foci by SPECT(AC), respectively. In 67 of 227 foci (29.5%), focus intensity/contrast increased after AC, whereas only 5 foci showed a decrease (P < 0.001). Sensitivity increased by 2.2% (P = 0.025; 95% CI: 0.02%-4.1%) as 5 foci were detected only by SPECT(AC). However, as these 3 patients were already diagnosed with systemic disease, there was no influence on the therapeutic strategy chosen.

Conclusion: Attenuation correction of somatostatin receptor scintigraphy-SPECT significantly improves focus visualization and, albeit slightly, also significantly increases sensitivity.

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Artifacts*
  • Female
  • Humans
  • Image Enhancement / methods*
  • Male
  • Middle Aged
  • Neuroendocrine Tumors / diagnosis*
  • Neuroendocrine Tumors / metabolism
  • Octreotide / analogs & derivatives*
  • Octreotide / pharmacokinetics
  • Radiation Dosage
  • Radiopharmaceuticals / pharmacokinetics
  • Receptors, Somatostatin / metabolism
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Subtraction Technique*
  • Systems Integration
  • Tomography, Emission-Computed, Single-Photon / methods*
  • Tomography, X-Ray Computed / methods*

Substances

  • Radiopharmaceuticals
  • Receptors, Somatostatin
  • indium-111-octreotide
  • Octreotide