Day-to-day test-retest variability of CBF, CMRO2, and OEF measurements using dynamic 15O PET studies

Mol Imaging Biol. 2011 Aug;13(4):759-68. doi: 10.1007/s11307-010-0382-1.

Abstract

Purpose: We assessed test-retest variability of cerebral blood flow (CBF), cerebral blood volume (CBV), cerebral metabolic rate of oxygen (CMRO(2)), and oxygen extraction fraction (OEF) measurements derived from dynamic (15)O positron emission tomography (PET) scans.

Procedures: In seven healthy volunteers, complete test-retest (15)O PET studies were obtained; test-retest variability and left-to-right ratios of CBF, CBV, OEF, and CMRO(2) in arterial flow territories were calculated.

Results: Whole-brain test-retest coefficients of variation for CBF, CBV, CMRO(2), and OEF were 8.8%, 13.8%, 5.3%, and 9.3%, respectively. Test-retest variability of CBV left-to-right ratios was <7.4% across all territories. Corresponding values for CBF, CMRO(2), and OEF were better, i.e., <4.5%, <4.0%, and <1.4%, respectively.

Conclusions: The test-retest variability of CMRO(2) measurements derived from dynamic (15)O PET scans is comparable to within-session test-retest variability derived from steady-state (15)O PET scans. Excellent regional test-retest variability was observed for CBF, CMRO(2), and OEF. Variability of absolute CBF and OEF measurements is probably affected by physiological day-to-day variability of CBF.

Publication types

  • Clinical Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aged, 80 and over
  • Blood Volume
  • Brain / metabolism*
  • Brain / physiopathology*
  • Brain Ischemia / diagnostic imaging
  • Brain Ischemia / pathology
  • Cerebrovascular Circulation / physiology*
  • Female
  • Humans
  • Male
  • Middle Aged
  • Oxygen / metabolism*
  • Oxygen Radioisotopes
  • Positron-Emission Tomography / methods*
  • Reproducibility of Results

Substances

  • Oxygen Radioisotopes
  • Oxygen