Coronary CT angiography of patients with a normal body mass index using 80 kVp versus 100 kVp: a prospective, multicenter, multivendor randomized trial

AJR Am J Roentgenol. 2011 Nov;197(5):W860-7. doi: 10.2214/AJR.11.6787.

Abstract

Objective: We determined the effect of reduced 80-kVp tube voltage on the radiation dose and image quality of coronary CT angiography (CTA) in patients with a normal body mass index (BMI).

Subjects and methods: A prospective, multicenter, multivendor trial was performed of 208 consecutive patients with a normal BMI (< 25 kg/m(2)) who had been referred for coronary CTA and did not have a history of coronary revascularization. Patients were randomized to 80-kVp imaging (n = 103) or 100-kVp imaging (n = 105). Three blinded readers graded interpretability and image quality. Study signal, noise, and contrast were also compared.

Results: Imaging with 80 kVp instead of 100 kVp was associated with 47% lower median radiation dose (median dose-length product, 62.0 mGy · cm [interquartile range, 54.0-123.3 mGy · cm] vs 117.0 mGy · cm [110.0-225.9 mGy · cm], respectively; 0.9 mSv [0.8-1.7 mSv] vs 1.6 mSv [1.4-3.2 mSv]; p < 0.001 for each) with no significant difference in interpretability (99% vs 99%; p = 0.99) or image quality (median score, 4.0 [interquartile range, 3.6-4.0] vs 4.0 [interquartile range, 3.8-4.0]; p = 0.20). Studies obtained using 80 kVp were associated with 27% increased signal (mean ± SD, 756 ± 157 vs 594 ± 105 HU; p < 0.001), 25% higher contrast (890 ± 156 vs 709 ± 108 HU; p < 0.001), and 50% greater noise (55 ± 15 vs 37 ± 12 HU; p < 0.001) with resultant 15% and 16% decreases in signal-to-noise (mean ± SD, 15 ± 5 vs 17 ± 5; p < 0.001) and contrast-to-noise (mean ± SD, 17 ± 6 vs 21 ± 5; p < 0.001) ratios, respectively.

Conclusion: Coronary CTA using 80 kVp instead of 100 kVp was associated with a nearly 50% reduction in radiation dose with no significant difference in interpretability and noninferior image quality despite lower signal-to-noise and contrast-to-noise ratios. The use of 80-kVp tube voltage should be considered in dose-reduction strategies for coronary CTA of individuals with a normal BMI.

Publication types

  • Multicenter Study
  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Body Mass Index*
  • Cardiac-Gated Imaging Techniques / methods*
  • Chi-Square Distribution
  • Comorbidity
  • Coronary Angiography / methods*
  • Coronary Disease / diagnostic imaging*
  • Female
  • Humans
  • Linear Models
  • Male
  • Middle Aged
  • Prospective Studies
  • Radiation Dosage
  • Radiographic Image Interpretation, Computer-Assisted
  • Statistics, Nonparametric
  • Tomography, X-Ray Computed / methods*