Acute clinical evaluation of a left ventricular automatic threshold determination algorithm based on evoked response sensing

Pacing Clin Electrophysiol. 2012 Mar;35(3):348-56. doi: 10.1111/j.1540-8159.2011.03287.x. Epub 2011 Dec 8.

Abstract

Introduction: Automatic pacing threshold (AT) testing may simplify device follow-up and improve device longevity. This study's objective was to evaluate the performance of a left ventricular (LV) evoked response sensing-based AT algorithm, for cardiac resynchronization therapy (CRT) devices.

Methods: Patients scheduled for CRT-D/P implant were enrolled. A manual step-down threshold test and a Left Ventricular Automatic Threshold (LVAT) test in each of four pacing vectors-LVTip→Can, LVTip→right ventricle (RV), = LVRing→Can, and LVRing→RV-were conducted. Patients were randomized to either 0.4-ms or 1.0-ms pacing pulse width and in the manual and LVAT test order. A blinded core lab electrophysiologist (EP) determined the threshold using the surface electrocardiogram (gold standard).

Results: Data from 70 patients were analyzed. Bipolar LV leads from three major manufacturers were used. A total of 273 AT tests were performed; 12 AT tests did not result in a threshold due to improper testing conditions. Of 261 eligible tests, 234 AT tests (89.6%) returned a threshold measurement. Of the 234 tests, in 233 tests (99.5%) the algorithm-determined threshold matched the EP-determined threshold for that test. A total of 16,689 capture and 526 noncapture beats were collected and the accuracy for detecting capture and noncapture were 98.5% and 99.7% with a two-sided 95% confidence level of (98.4%, 98.7%) and (99.4%, 100%), respectively. No AT threshold measurement was lower than the EP-determined threshold.

Conclusion: In this study, the results suggest that the LVAT algorithm is accurate at determining pacing thresholds in multiple pacing configurations and a wide range of LV leads in CRT-D/P patients.

Publication types

  • Multicenter Study
  • Randomized Controlled Trial

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Algorithms*
  • Cardiac Pacing, Artificial / methods*
  • Cardiac Resynchronization Therapy Devices*
  • Electrocardiography / methods
  • Female
  • Heart Ventricles / physiopathology*
  • Humans
  • Male
  • Middle Aged