Instrumentation to record evoked potentials for closed-loop control of deep brain stimulation

Annu Int Conf IEEE Eng Med Biol Soc. 2011:2011:6777-80. doi: 10.1109/IEMBS.2011.6091671.

Abstract

Closed-loop deep brain stimulation (DBS) systems offer promise in relieving the clinical burden of stimulus parameter selection and improving treatment outcomes. In such a system, a feedback signal is used to adjust automatically stimulation parameters and optimize the efficacy of stimulation. We explored the feasibility of recording electrically evoked compound action potentials (ECAPs) during DBS for use as a feedback control signal. A novel instrumentation system was developed to suppress the stimulus artifact and amplify the small magnitude, short latency ECAP response during DBS with clinically relevant parameters. In vitro testing demonstrated the capabilities to increase the gain by a factor of 1,000× over a conventional amplifier without saturation, reduce distortion of mock ECAP signals, and make high fidelity recordings of mock ECAPs at latencies of only 0.5 ms following DBS pulses of 50 to 100 μs duration. Subsequently, the instrumentation was used to make in vivo recordings of ECAPs during thalamic DBS in cats, without contamination by the stimulus artifact. The signal characteristics were similar across three experiments, suggesting common neural activation patterns. The ECAP recordings enabled with this novel instrumentation may provide insight into the type and spatial extent of neural elements activated during DBS, and could serve as feedback control signals for closed-loop systems.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Amplifiers, Electronic
  • Animals
  • Artifacts
  • Cats
  • Deep Brain Stimulation / instrumentation*
  • Deep Brain Stimulation / methods*
  • Electric Stimulation
  • Electrodes
  • Equipment Design
  • Evoked Potentials
  • Feedback
  • Signal Processing, Computer-Assisted
  • Thalamus / pathology