Standardized uptake values for [¹⁸F] FDG in normal organ tissues: comparison of whole-body PET/CT and PET/MRI

Eur J Radiol. 2013 May;82(5):870-6. doi: 10.1016/j.ejrad.2013.01.008. Epub 2013 Feb 8.

Abstract

Purpose: To compare maximum and mean standardized uptake values (SUVmax/mean) of normal organ tissues derived from [(18)F]-fluoro-desoxyglucose (FDG) positron emission tomography/magnetic resonance imaging (PET/MRI) using MR attenuation correction (MRAC) (DIXON-based 4-segment μ-map) with [(18)F]-FDG positron emission tomography/computed tomography (PET/CT) using CT-based attenuation correction (CTAC).

Methods and materials: In 25 oncologic patients (15 men, 10 women; age 57 ± 13 years) after routine whole-body FDG-PET/CT (60 min after injection of 290 ± 40 MBq [(18)F]-FDG) a whole-body PET/MRI was performed (Magnetom Biograph mMR, Siemens Healthcare, Erlangen, Germany). Volumes of interest of 1.0 cm(3) were drawn in 7 physiological organ sites in MRAC-PET and the corresponding CTAC-PET images manually. Spearman correlation coefficients were calculated to compare MRAC- and CTAC based SUV values; Wilcoxon-Matched-Pairs signed ranks test was performed to test for potential differences.

Results: The mean delay between FDG-PET/CT and PET/MRI was 92 ± 18 min. Excellent correlations of SUV values were found for the heart muscle (SUVmax/mean: R=0.97/0.97); reasonably good correlations were found for the liver (R=0.65/0.72), bone marrow (R=0.42/0.41) and the SUVmax of the psoas muscle (R=0.41). For subcutaneous fat, the correlation coefficient was 0.66 for SUVmean (p<0.05). Correlations between MRAC and CTAC were non-significant for SUVmean of the psoas muscle, SUVmax of subcutaneous fat, SUVmax and SUVmean of the lungs, SUVmax and SUVmean of the blood-pool. The median SUVmax and SUVmean in MRAC-PET were lower than the respective CTAC values in all organs (p<0.05) but heart (SUVmax) and the bone marrow (SUVmean).

Conclusion: In conclusion, in oncologic patients examined with PET/CT and PET/MRI SUVmax and SUVmean values generally correlate well in normal organ tissues, except the lung, subcutaneous fat and the blood pool. SUVmax and SUVmean derived from PET/MRI can be used reliably in clinical routine.

Publication types

  • Comparative Study

MeSH terms

  • Female
  • Fluorodeoxyglucose F18 / pharmacokinetics*
  • Germany
  • Humans
  • Magnetic Resonance Imaging / methods*
  • Magnetic Resonance Imaging / standards
  • Male
  • Metabolic Clearance Rate
  • Middle Aged
  • Multimodal Imaging / methods*
  • Multimodal Imaging / standards
  • Organ Specificity
  • Positron-Emission Tomography*
  • Radiopharmaceuticals / pharmacokinetics
  • Reference Values
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Tissue Distribution
  • Tomography, X-Ray Computed*
  • Whole Body Imaging / methods*
  • Whole Body Imaging / standards

Substances

  • Radiopharmaceuticals
  • Fluorodeoxyglucose F18