Assessment of IcareONE rebound tonometer for self-measuring intraocular pressure

Acta Ophthalmol. 2014 May;92(3):243-8. doi: 10.1111/aos.12108. Epub 2013 Apr 16.

Abstract

Purpose: To evaluate the precision of the IcareONE rebound tonometer, which was developed for self-measuring intraocular pressure (IOP) and to compare IcareONE measurement with Goldmann applanation tonometry (GAT).

Methods: Twenty-four healthy eyes and 81 glaucomatous eyes were enrolled. IOP measurements (three times per session) with IcareONE were made in a random order by an ophthalmologist (Icare(O)) and by the subject (Icare(S)). Intraclass correlation coefficients (CCs), kappa values and mean values of IOP were compared among the two types of Icare recordings and GAT. Bland-Altman analysis was used to assess agreement between methods. Multiple regression analysis was performed to identify the subject factors that influenced the discordant measurements between IcareONE and GAT.

Results: The mean value of Icare(O) and Icare(S) measurements was 13.5±5.2 and 13.5±5.4 mmHg, respectively, neither of which was significantly different from GAT (13.8±4.4). The intrarater CC of Icare(O) and Icare(S) was 0.968 and 0.885, respectively. The intermethod CC and weighted kappa between Icare(O) and Icare(S) were 0.907 and 0.684, respectively. All pairwise correlations between the two types of IOP measurement showed coefficients of determination >0.8. Bland-Altman analysis did not show any proportional biases. Multiple regression analysis revealed that the differences between GAT and Icare(O) or Icare(S) were positively correlated with central corneal thickness (CCT) and negatively correlated with age.

Conclusions: Intraocular pressure measurements with IcareONE by a physician and by the subject showed excellent agreement with GAT measurements; IcareONE measurements between a physician and the subject had high intrarater reliability, and good agreement thicker CCT led IcareONE measurement to overestimate IOP, while higher age caused it to underestimate IOP compared with GAT.

Keywords: IcareONE; intraocular pressure; rebound tonometer; self-measurements.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Cross-Sectional Studies
  • Equipment Design
  • Female
  • Follow-Up Studies
  • Glaucoma / diagnosis*
  • Glaucoma / physiopathology
  • Humans
  • Intraocular Pressure*
  • Male
  • Middle Aged
  • Ocular Hypertension / diagnosis*
  • Ocular Hypertension / physiopathology
  • Prospective Studies
  • Reproducibility of Results
  • Self Care / methods*
  • Tonometry, Ocular / instrumentation*
  • Young Adult