Use of flow-diverting devices in fusiform vertebrobasilar giant aneurysms: a report on periprocedural course and long-term follow-up

AJNR Am J Neuroradiol. 2014 Jul;35(7):1346-52. doi: 10.3174/ajnr.A3859. Epub 2014 Feb 27.

Abstract

Background and purpose: Fusiform vertebrobasilar giant aneurysms are a rare (<1% of all intracranial aneurysms) but challenging aneurysm subtype. Little data are available on the natural history of this aneurysm subtype and the impact of the use of flow-diverting stents on the long-term clinical and imaging follow-up. In this article, we present our experience with the treatment of fusiform vertebrobasilar giant aneurysms by flow diverting stents. We aim to stimulate a discussion of the best management paradigm for this challenging aneurysm subtype.

Materials and methods: We retrospectively identified 6 patients with fusiform vertebrobasilar giant aneurysms who had been treated with flow-diverting stents between October 2009 and March 2012 in our center. The available data were re-evaluated. The modified Rankin Scale score was assessed before intervention, during the stay in hospital, and at discharge.

Results: Six patients were identified (all male; age range, 49-71 years; median age, 60 years). Handling of material was successful in all cases. No primary periprocedural complications occurred. The mean follow-up was 13 months (15 days to 29 months). During follow-up, 3 of 6 patients had recurrent cerebral infarctions, but no patient experienced SAH. Two patients presented with acute thrombotic stent occlusion. The modified Rankin Scale score was not higher than 3 in any of the cases before intervention, whereas the best mRS score at the last follow-up was 5. Four of 6 patients died during follow-up.

Conclusions: Endovascular treatment of fusiform vertebrobasilar giant aneurysms with flow-diverting devices is feasible from a technical point of view; however, changes in hemodynamics with secondary thrombosis are not predictable. We currently do not intend to treat fusiform vertebrobasilar giant aneurysms with flow-diverting devices until we have further understanding of the pathophysiology, natural history, and hemodynamic effects of flow diversion.

Publication types

  • Case Reports

MeSH terms

  • Aged
  • Cerebral Revascularization / instrumentation*
  • Endovascular Procedures / instrumentation*
  • Equipment Failure Analysis
  • Female
  • Humans
  • Intracranial Aneurysm / diagnostic imaging*
  • Intracranial Aneurysm / surgery*
  • Longitudinal Studies
  • Male
  • Middle Aged
  • Prosthesis Design
  • Radiography
  • Stents*
  • Treatment Outcome
  • Vertebrobasilar Insufficiency / diagnostic imaging*
  • Vertebrobasilar Insufficiency / surgery*