Comparison of the Roche COBAS AmpliPrep/COBAS TaqMan HIV-1 test v1.0 with v2.0 in HIV-1 viral load quantification

Kaohsiung J Med Sci. 2015 Apr;31(4):188-93. doi: 10.1016/j.kjms.2014.12.010. Epub 2015 Feb 3.

Abstract

Roche modified the COBAS AmpliPrep/COBAS TaqMan human immunodeficiency virus type 1 (HIV-1) test version 1.0 (CAP/CTM v1.0), resulting in the COBAS AmpliPrep/COBAS TaqMan HIV-1 test version 2.0 (CAP/CTM v2.0). The aim of this study was to evaluate the performance of the CAP/CTM v2.0 and to compare this performance with that of the CAP/CTM v1.0. The study was conducted in a small local study group in Kaohsiung Medical University Hospital, Kaohsiung, Taiwan. A total of 86 plasma samples from HIV-1-seropositive patients were tested using the two assays. The correlation and concordance of results between the two assays were calculated. The CAP/CTM v2.0 generated higher values than did the CAP/CTM v1.0, and five samples (5.8%) yielded a difference of > 1 log10 copies/mL. In addition, our data show that CAP/CTM v1.0 and CAP/CTM v2.0 yielded relatively consistent values for 23 samples with low viral loads (< 200 copies/mL). Furthermore, when viral loads were in a medium range (2-5 log10 copies/mL), the results of the two assays were more compatible. This study shows a good correlation between CAP/CTM v1.0 and v2.0 in HIV-1 viral load measurement. Further attention must be paid to those cases in which measured viral loads present larger differences between the two assays.

Keywords: CAP/CTM v1.0; CAP/CTM v2.0; HIV-1; Viral loads.

MeSH terms

  • Adult
  • Aged
  • Female
  • HIV Infections / diagnosis*
  • HIV Infections / virology*
  • HIV-1 / genetics
  • HIV-1 / pathogenicity*
  • Humans
  • Male
  • RNA, Viral / genetics
  • Viral Load

Substances

  • RNA, Viral