Use of a Combined Blood-Glucose- and ß-Ketone-Measuring Device Improves Glycemic Control in Insulin-Treated Patients With Diabetes: The Gold Plus Study

J Diabetes Sci Technol. 2015 May 18;9(6):1270-4. doi: 10.1177/1932296815587936.

Abstract

Background: Using self-measurements of blood glucose (SMBG) is daily routine for patients with insulin-treated diabetes, however measuring ß-ketones in blood is not widespread. How the use of a combined device, which can measure both, is accepted in daily routine by patients and will lead to better glycemic control is not well studied.

Methods: This multicenter, prospective, noninterventional study assessed the impact of routine use of the GlucoMen® LX Plus on patient acceptance, usage and glycemic control among insulin-treated patients with diabetes mellitus type 1 and type 2. A1c and self-reported frequency of SMBG were evaluated at baseline and also postprandial SMBG, ß-ketone measurements, and use of reminders after 3 and 6 months of use. A total of 631 patients, 254 type 1/350 type 2 (27 no type specified), with mean (SD) baseline A1c 8.5% (1.5), age 54.6 (15.6) years, and 47.3% female were studied.

Results: Frequent use of SMBG at baseline led to a higher decrease in A1c at 6 month (V3): -0.3% if SMBG measured up to 1/day versus -0.9% in 4-6/day. Increase of SMBG frequency during the study showed also a negative correlation to A1c, 9.2% at V1 versus 7.6% at V3. Postprandial SMBG was done by 77.7% and ß-ketone measurements by 45.5% of all patients; the reminders were used by 33.4% and led to an increased frequency of SMBG at 6 months.

Conclusions: A combined device for SMBG and ß-ketone measurements is well accepted by patients with insulin-treated diabetes and can lead through the avoidance or detection of ketoacidosis/increased frequency of SMBG and increased awareness of the patients to an improved glycemic outcome.

Keywords: blood glucose meter; blood ketone meter; self-monitoring of blood glucose; self-monitoring of blood ketones.

Publication types

  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Biomarkers / blood
  • Blood Glucose / drug effects
  • Blood Glucose / metabolism*
  • Blood Glucose Self-Monitoring / instrumentation*
  • Diabetes Mellitus, Type 1 / blood
  • Diabetes Mellitus, Type 1 / diagnosis*
  • Diabetes Mellitus, Type 1 / drug therapy
  • Diabetes Mellitus, Type 2 / blood
  • Diabetes Mellitus, Type 2 / diagnosis*
  • Diabetes Mellitus, Type 2 / drug therapy
  • Equipment Design
  • Female
  • Germany
  • Glycated Hemoglobin / metabolism
  • Humans
  • Hypoglycemic Agents / therapeutic use
  • Insulin / therapeutic use
  • Ketones / blood*
  • Male
  • Middle Aged
  • Patient Satisfaction
  • Predictive Value of Tests
  • Prospective Studies
  • Time Factors
  • Treatment Outcome

Substances

  • Biomarkers
  • Blood Glucose
  • Glycated Hemoglobin A
  • Hypoglycemic Agents
  • Insulin
  • Ketones
  • hemoglobin A1c protein, human