Long-term (8-10 years) outcomes after biodegradable polymer-coated biolimus-eluting stent implantation

Heart. 2017 Jul;103(13):1002-1008. doi: 10.1136/heartjnl-2016-310591. Epub 2017 Jan 17.

Abstract

Objective: Efficacy and safety data on biodegradable polymer-coated biolimus-eluting stent (BP-BES) are currently limited to 5 years. We evaluated longer term (8-10 years) clinical and angiographic outcomes after BP-BES implantation.

Methods: Between 2005 and 2008, 243 patients (301 lesions) underwent BP-BES implantation. The primary clinical outcome measure was defined as any target lesion revascularisation (TLR). Absolute serial angiographic studies without any concomitant TLR within 2 years after the procedure were performed in 55 patients (65 lesions) at postprocedure, mid-term (within 1 year), late term (between 1 and 2 years) and very late term (beyond 2 years).

Results: The median follow-up duration was 9.4 years (IQR 8.2-10.2 years). The 8-year cumulative incidence of any TLR was 20.3%. The increase rate was approximately 7% per year in the first 2 years, but decelerated to approximately 1.2% per year beyond 2 years after the procedure. The minimal lumen diameter significantly decreased from postprocedure (2.63±0.44 mm) to mid-term (2.43±0.59 mm, p=0.002) and from late term (2.27±0.63 mm) to very late term (1.98±0.73 mm, p=0.002). The 8-year cumulative incidences of definite or probable stent thrombosis (ST) and major bleeding (Bleeding Academic Research Consortium (BARC) ≥3) were 0.5% and 12.0%, respectively. Definite ST was none within 10 years in the entire cohort.

Conclusions: The long-term clinical outcomes after BP-BES implantation were favourable, although angiographic late progression of luminal narrowing did not reach a plateau. The incidence of ST remained notably low, whereas that of major bleeding gradually increased.

MeSH terms

  • Absorbable Implants*
  • Aged
  • Coated Materials, Biocompatible*
  • Coronary Angiography
  • Coronary Artery Disease / diagnosis
  • Coronary Artery Disease / surgery*
  • Drug-Eluting Stents*
  • Female
  • Follow-Up Studies
  • Forecasting*
  • Humans
  • Immunosuppressive Agents / pharmacology
  • Male
  • Myocardial Revascularization / methods
  • Polymers*
  • Prosthesis Design
  • Retrospective Studies
  • Sirolimus / analogs & derivatives*
  • Sirolimus / pharmacology
  • Treatment Outcome

Substances

  • Coated Materials, Biocompatible
  • Immunosuppressive Agents
  • Polymers
  • umirolimus
  • Sirolimus