Validation of the G.LAB MD2200 wrist blood pressure monitor according to the European Society of Hypertension, the British Hypertension Society, and the International Organization for Standardization Protocols

Blood Press Monit. 2017 Apr;22(2):101-104. doi: 10.1097/MBP.0000000000000234.

Abstract

Objective: To validate the G.LAB MD2200 automated wrist blood pressure (BP) monitors according to the European Society of Hypertension International Protocol (ESH-IP) revision 2010, the British Hypertension Society (BHS), and the International Organization for Standardization (ISO) 81060-2:2013 protocols.

Materials and methods: The device was assessed on 33 participants according to the ESH requirements and was then tested on 85 participants according to the BHS and ISO 81060-2:2013 criteria. The validation procedures and data analysis followed the protocols precisely.

Results: The G.LAB MD2200 devices passed all parts of ESH-IP revision 2010 for both systolic and diastolic BP, with a device-observer difference of 2.15±5.51 and 1.51±5.16 mmHg, respectively. The device achieved A/A grading for the BHS protocol and it also fulfilled the criteria of ISO 81060-2:2013, with mean differences of systolic and diastolic BP between the device and the observer of 2.19±5.21 and 2.11±4.70 mmHg, respectively.

Conclusion: The G.LAB MD2200 automated wrist BP monitor passed the ESH-IP revision 2010 and the ISO 81060-2:2013 protocol, and achieved the A/A grade of the BHS protocol, which can be recommended for self-measurement in the general population.

Publication types

  • Clinical Trial
  • Validation Study

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Blood Pressure Determination / instrumentation*
  • Blood Pressure Determination / methods
  • Blood Pressure Determination / standards*
  • Blood Pressure Monitors / standards*
  • Blood Pressure*
  • Humans
  • Hypertension / physiopathology*
  • Middle Aged
  • Societies, Medical