In Vivo Differentiation of Uric Acid Versus Non-Uric Acid Urinary Calculi With Third-Generation Dual-Source Dual-Energy CT at Reduced Radiation Dose

AJR Am J Roentgenol. 2018 Feb;210(2):358-363. doi: 10.2214/AJR.17.18091. Epub 2017 Nov 22.

Abstract

Objective: The objective of our study was to evaluate in vivo urinary calculus characterization with third-generation dual-source dual-energy CT (DECT) at reduced versus standard radiation dose.

Subjects and methods: One hundred fifty-three patients requiring unenhanced CT for suspected or known urolithiasis were prospectively included in our study. They underwent two acquisitions at reduced-dose CT (90 kV and 50 mAsref; Sn150 kV and 31 mAsref, where Sn denotes the interposition of a tin filter in the high-energy beam) and standard-dose CT (90 kV and 50 mAsref; Sn150 kV and 94 mAsref). One radiologist interpreted the reduced-dose examinations before the standard-dose examinations during the same session. Among 103 patients (23 women, 80 men; mean age ± SD, 50 ± 15 years; age range, 18-82 years) with urolithiasis, dedicated DECT software measured the maximal diameter and CT numbers, calculated the DECT number ratio, and labeled with a color code each calculus visualized by the radiologist as uric acid (UA) or non-UA. Volume CT dose index (CTDIvol) and dose-length product (DLP) were recorded.

Results: The radiologist visualized 279 calculi on standard-dose CT and 262 on reduced-dose CT; 17 calculi were missed on reduced-dose CT, all of which were ≤ 3 mm. Among the 262 calculi visualized at both doses, the CT number ratio was obtained with the software for 227 calculi and was not different between the doses (p = 0.093). Among these 262 calculi, 197 were labeled at both doses; 194 of the 197 labeled calculi were labeled with the same color code. Among the 65 remaining calculi, 48 and 61 (all ≤ 5 mm) were not labeled at standard-dose and reduced-dose CT (p = 0.005), respectively. At reduced-dose CT, the mean CTDIvol was 2.67 mGy and the mean DLP was 102.2 mGy × cm.

Conclusion: With third-generation dual-source DECT, a larger proportion of calculi ≤ 5 mm are not characterized as UA or non-UA at a reduced dose.

Keywords: CT number ratio; dual-energy CT; radiation dose; uric acid; urinary calculi.

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Aged, 80 and over
  • Diagnosis, Differential
  • Female
  • Humans
  • Male
  • Middle Aged
  • Prospective Studies
  • Radiation Dosage
  • Radiographic Image Interpretation, Computer-Assisted
  • Radiography, Dual-Energy Scanned Projection*
  • Sensitivity and Specificity
  • Software
  • Tomography, X-Ray Computed / methods*
  • Uric Acid
  • Urinary Calculi / chemistry
  • Urinary Calculi / diagnostic imaging*

Substances

  • Uric Acid