Intracardiac echocardiography versus fluoroscopy for endovascular and endocardial catheter navigation during cryo-ablation of the slow pathway in AVNRT patients

Cardiovasc Ultrasound. 2019 Jun 11;17(1):12. doi: 10.1186/s12947-019-0162-2.

Abstract

Background: A new zero-fluoroscopy technique for electrophysiology catheter navigation relying on intracardiac echocardiography (ICE) has been recently reported (Ice&ICE trial). We investigated potential differences in efficacy, safety or procedural performance between conventional fluoroscopy- and ICE-guided cryothermal ablation (CA) in symptomatic AVNRT patients.

Methods: Clinical and electrophysiological data of AVNRT patients included in the Ice&ICE trial (22 patients, 16 females; =zero-fluoroscopy group) were compared to those of consecutive AVNRT patients, who underwent fluoroscopy-guided CA (25 patients, 17 females; = fluoroscopy group) during the last 2 years in our institution.

Results: Slow pathway ablation or modulation was successful in all patients. Fluoroscopy time and radiation dose in the fluoroscopy group were 11.2 ± 9.0 min and 20.3 ± 16.2Gycm2, whereas no fluoroscopy was used in the opposite group (p < 0.001, respectively). EPS duration was not different between the groups (zero-fluoroscopy:101.6 ± 40.2 min, fluoroscopy:99.4 ± 37.2 min, p = n.s.). Catheter placement time was significantly shorter in the fluoroscopy group (2.2 ± 1.6 min vs. 12.0 ± 7.5 min, p < 0.05), whereas cryo-application duration (from the first cryo-mapping to the last CA) was significantly shorter in the zero-fluoroscopy group (27.5 ± 37.0 min vs. 38.1 ± 33.9 min, p < 0.05). Mean cryo-mapping and CA applications were numerically lower in the zero-fluoroscopy group (CM:7.5 ± 5.7 vs. 8.8 ± 6.2; CA:3.1 ± 1.7 vs. 3.2 ± 2.0, p = n.s.). No major adverse events occurred in both groups. After 15.0 ± 4.2 months, arrhythmia recurrence was not different between the groups (4.5% vs. 8.0%, p = n.s.).

Conclusions: Zero-fluoroscopy ICE-guided EP catheter navigation shows comparable efficacy and safety to fluoroscopic guidance during CA in AVNRT patients. ICE visualization of catheters and endocardial structures within the triangle of Koch shortens the cryo-application duration, though time needed for catheter placement is longer, when compared with conventional fluoroscopic guidance, which results in similar mean EPS duration with both navigation techniques.

Trial registration: (German Clinical Trials Register ID: DRKS00011360 ; Registration Date 14.12.2016).

Keywords: AVNRT; Catheter ablation; Cryo-ablation; Intracardiac echocardiography; Zero-fluoroscopy.

MeSH terms

  • Cardiac Catheterization / methods*
  • Cryosurgery / methods*
  • Echocardiography / methods*
  • Electrocardiography
  • Endovascular Procedures / methods*
  • Feasibility Studies
  • Female
  • Fluoroscopy / methods*
  • Humans
  • Imaging, Three-Dimensional
  • Male
  • Middle Aged
  • Pericardium
  • Reproducibility of Results
  • Retrospective Studies
  • Surgery, Computer-Assisted / methods*
  • Tachycardia, Atrioventricular Nodal Reentry / diagnosis
  • Tachycardia, Atrioventricular Nodal Reentry / physiopathology
  • Tachycardia, Atrioventricular Nodal Reentry / surgery*
  • Ultrasonography, Interventional / methods